To measure the balance of dental occlusion, the patient must perform two clenching tests of five seconds each - the first with cotton rolls, the second one without. Once these trials are done, the report is automatically generated.
- To proceed with a new acquisition, from the Home Screen of the software, select the desired patient. Then, click on Occlusal Test
- Insert the cotton rolls between the arches on the second premolars and molars
- Ask the patient to clench their teeth as hard as possible
- To start the Calibration, press the Rec button
. The recording will stop automatically after five seconds
⚠️🚨Teethan's Calibration Test is mandatory and it is meant to adapt the technology to each and every patient's neuromuscular strategy of occlusion.
- At the end of the Calibration, you will see - on the left -its results. To perform multiple Calibrations, just press Rec again.
- Once you have acquired the Occlusal Calibration Test, click on
to move on to the final part of the Test.
- After removing the cotton rolls, ask the patient to clench as hard as possible to allow dental intercuspation. To start the recording, press the Rec
button

During the test execution, the electrical activity of the analyzed muscles is displayed in real-time, using bars that get filled with color proportionally to the degree of muscular activation. Each of the four muscles is identified by a characteristic color:
Right Masseter:
Left Masseter:
Right Temporalis:
Left Temporalis: